Video Versus Direct Laryngoscopy and The Hemodynamic Response and Myocardial Workload During Tracheal Intubation Under Standardized Anesthetic Induction: A Randomized Trial

Authors

  • Imam Safi'i Department of Anesthesiology and Intensive Care, Faculty of Medicine, Universitas Brawijaya/Dr. Saiful Anwar Regional General Hospital, Malang, Indonesia
  • Arie Zainul Fatoni Department of Anesthesiology and Intensive Care, Faculty of Medicine, Universitas Brawijaya/Dr. Saiful Anwar Regional General Hospital, Malang, Indonesia
  • Taufiq Agus Siswagama Department of Anesthesiology and Intensive Care, Faculty of Medicine, Universitas Brawijaya/Dr. Saiful Anwar Regional General Hospital, Malang, Indonesia
  • Ahmad Feza Fadhlurrahman Department of Anesthesiology and Intensive Care, Faculty of Medicine, Universitas Brawijaya/Dr. Saiful Anwar Regional General Hospital, Malang, Indonesia

DOI:

https://doi.org/10.37275/jacr.v7i1.963

Keywords:

Airway management, Direct laryngoscopy, Hemodynamics, Rate pressure product, Sympathoadrenal response, Video laryngoscopy

Abstract

Background. Tracheal intubation triggers a sympathoadrenal reflex that raises heart rate and blood pressure. Whether video laryngoscopy (VL) reduces this response is debated.

Objective. To compare GlideScope VL with Macintosh direct laryngoscopy (DL) for the hemodynamic response and rate pressure product (RPP) under standardized induction.

Methods. In this single-blind randomized trial, 40 adults (ASA I–II) received VL (n = 20) or DL (n = 20). Induction was fentanyl 2 mcg/kg, propofol 2 mg/kg, atracurium 0.5 mg/kg, and 1.5% sevoflurane, with laryngoscopy 3 minutes after the relaxant. Heart rate (HR), systolic (SBP) and mean arterial pressure (MAP) were recorded at baseline (T0) and 1 (T1), 2 (T2), and 5 (T5) minutes after intubation.

Results. HR was lower with VL at T1 (75.45 ± 11.23 vs. 90.15 ± 15.22 bpm, p = 0.001) and T2 (74.15 ± 10.80 vs. 89.85 ± 14.50 bpm, p < 0.001). MAP showed a Time × Group interaction (p < 0.001) and was lower with VL at T1 (96.30 ± 7.80 vs. 108.40 ± 9.60 mmHg, p < 0.001). SBP at T1 did not differ significantly (p = 0.052). RPP was lower with VL at T1 (8,952 ± 1,480 vs. 11,453 ± 2,210) and T2 (8,649 ± 1,390 vs. 11,213 ± 2,150 bpm·mmHg; both p < 0.001). First-pass success was 100% vs. 90% (p = 0.487).

Conclusion. In young ASA I–II patients, VL was associated with lower HR, MAP, and RPP 1–2 minutes after intubation than DL; SBP and first-pass success did not differ significantly.

Authors

  • Imam Safi'i1*
  • Arie Zainul Fatoni1
  • Taufiq Agus Siswagama1
  • Ahmad Feza Fadhlurrahman1
  1. 1Department of Anesthesiology and Intensive Care, Faculty of Medicine, Universitas Brawijaya/Dr. Saiful Anwar Regional General Hospital, Malang, Indonesia

Corresponding author Imam Safi'i — dr.imams85.tnial@gmail.com

Downloads

Published

2026-05-29

How to Cite

Video Versus Direct Laryngoscopy and The Hemodynamic Response and Myocardial Workload During Tracheal Intubation Under Standardized Anesthetic Induction: A Randomized Trial. (2026). Journal of Anesthesiology and Clinical Research, 7(1), 41-49. https://doi.org/10.37275/jacr.v7i1.963

Article Metrics

11Abstract views
5Downloads

Views and downloads recorded by this journal.

Most read articles by the same author(s)